Temperature Measurement for Validation of Clinical Moist Sterilizers
기관명 | NDSL |
---|---|
저널명 | 中華醫院感染學雜志 = Chinese journal of nosocomiology |
ISSN | 1005-4529, |
ISBN |
저자(한글) | ZHANG, Qing,ZHANG, Jintao,GAO, Hong,DUAN, Yuning,QIU, Ping |
---|---|
저자(영문) | |
소속기관 | |
소속기관(영문) | |
출판인 | |
간행물 번호 | |
발행연도 | 2008-01-01 |
초록 | OBJECTIVE To report the investigation on validating clinical moist sterilizer by applying wireless thermo-loggers in accordance with the advanced Euro and British standards.METHODS The validation had been implemented with microbiological tests and thermometric tests to measure the temperature distribution of the chamber of sterilizer,the temperature and time relation inside the tested package of small and challenge loads,the pressure of the chamber and the temperature beside the drainage.RESULTS The measurements presented the relation between temperature and time measurements of the spots inside the chamber of the tested sterilization loads.CONCLUSIONS The measurement results indicate directly the operation condition and the temperature-time relation inside sterilization loads.These measurements are helpful for controlling sterilization quality.The results of microbiological tests are negative,which are consistent with that of thermomeric tests. |
원문URL | http://click.ndsl.kr/servlet/OpenAPIDetailView?keyValue=03553784&target=NART&cn=NART51531049 |
첨부파일 |
과학기술표준분류 | |
---|---|
ICT 기술분류 | |
DDC 분류 | |
주제어 (키워드) | Sterilization,Validation,Sterilization load |