Aptamer-based Piezoelectric Quartz Crystal Biosensor
기관명 | NDSL |
---|---|
저널명 | 中華醫院感染學雜志 = Chinese journal of nosocomiology |
ISSN | 1005-4529, |
ISBN |
저자(한글) | QI, Yongzhi,YAO, Chunyan,LUO, Yang,FU, Weiling |
---|---|
저자(영문) | |
소속기관 | |
소속기관(영문) | |
출판인 | |
간행물 번호 | |
발행연도 | 2008-01-01 |
초록 | OBJECTIVE To establish an aptamer-based piezoelectric .sensor assay to detect human immunoglobulin E (IgE) and acquire the parameters about this kind of biosensor. METHODS Anti-IgE aptamers were immobilized on the gold surface of piezoelectric quartz crystal sensor with biotin-avidin method. The reaction between anti-IgE aptamers and IgE was monitored in time on piezoelectric sensor. Different concentrations of IgE were detected and the calibration curve of IgE was drawn. Non-specific response signals of BSA, IgG and IgE were monitored to show the detecting specificity of aptamer-based sensor. RESULTS For the aptamer-based sensor, the lowest limit of detection was 0. 055mg/L and the linear range was 0.1-2. 5mg/L. Non-specific signals of this sensor were less than 5M of specific signals of IgE at the same concentration. CONCLUSIONS The aptamer-based piezoelectric sensor can detect 5.5 ng IgE while the non-specific signals are very low. So this aptamer-based sensor is hopeful to be applied to clinical laboratory diagnosis. |
원문URL | http://click.ndsl.kr/servlet/OpenAPIDetailView?keyValue=03553784&target=NART&cn=NART51530989 |
첨부파일 |
과학기술표준분류 | |
---|---|
ICT 기술분류 | |
DDC 분류 | |
주제어 (키워드) | QCM,Sensor,Aptamer,Probe |