고분해능삼중사중극자 액체크로마토그래피 텐덤질량분석기
| 기관명 | ZEUS |
|---|---|
| 장비번호 | |
| 제작사 | Thermo Scientific |
| 모델명 | TSQ Quantum Ultra EMR |
| 장비사양 | |
| 취득일자 | 2007-01-03 |
| 취득금액 |
| 보유기관명 | 한국표준과학연구원 |
|---|---|
| 보유기관코드 | |
| 활용범위 | |
| 활용상태 | |
| 표준코드 | B602 |
| 표준분류명 | 시험 |
| 시설장비 설명 | C. 성능 1. MS/MS function with a standard mass range of 30 to 1,250 amu or more 2. High resolution precursor ion selection of less or equal then 0.1 Dalton. 3. Mass Stability with mass assignment within +/- 0.050 Da over a 24 hour period 4. Mass Resolution continuously adjustable to better than 0.1 Da peak width (FWHM) across the entire mass range. 5. Sensitivity of ESI: A 5ul injection(1 pg) of a 200 fg/ul(0.329 fmol/ul) reserpine solution at a flow rate of 400ul/min will produce a minimum signal to noise ratio of 50:1 for the transition of the protonated molecular ion at m/z 609.3 to the fragment ion at m/z 195.1 in selected reaction monitoring(SRM) mode with Q1 and Q3 resolution set to 0.7 Da FWHM. or better A 5ul injection(1 pg) of a 200 fg/ul(0.329 fmol/ul) reserpine solution at a flow rate of 400ul/min will produce a minimum signal to noise ratio of 50:1 for the transition of the protonated molecular ion at m/z 609.3 to the fragment ion at m/z 195.1 in selected reaction monitoring(SRM) mode with Q |
| 장비이미지코드 | http://www.zeus.go.kr/storage/images//equip/photo/201712/20161007100736_20110125000000100357 NFEC-2011-01-139096.jpg |
| 장비위치주소 | 한국표준과학연구원 인증표준물질(CRM)동 |
| NFEC 등록번호 | NFEC-2011-01-139096 |
| 예약방법 | |
| 카타로그 URL | |
| 메뉴얼 URL | |
| 원문 URL | http://www.zeus.go.kr/resv/equip/read/Z-NTIS-0049629 |
| 첨부파일 |
| 과학기술표준분류 | |
|---|---|
| ICT 기술분류 | |
| 주제어 (키워드) |