전자선형가속기
| 기관명 | ZEUS |
|---|---|
| 장비번호 | |
| 제작사 | Elekta |
| 모델명 | Synergy® Platform |
| 장비사양 | |
| 취득일자 | 2016-02-24 |
| 취득금액 |
| 보유기관명 | 한국표준과학연구원 |
|---|---|
| 보유기관코드 | |
| 활용범위 | |
| 활용상태 | |
| 표준코드 | F0 |
| 표준분류명 | |
| 시설장비 설명 | □ Reproducibility - Maximum coefficient of variation of the ratio of number of dose monitor units and absorbed dose is ≤0.5%. This is applicable for any dose rate and to every individual segment of all conformal and intensity modulated delivery techniques. □ Proportionality (linearity) - The maximum deviation of measured absorbed dose by an external dose meter at the isocenter from the number of delivered monitor units multiplied by the proportionality (calibration) factor is ≤1% or 0.1 MU, whichever is greater for 1 to 999 MU. □ MLCi2 Head - Lower Leakage Precision Multileaf Collimator - The MLCi2 Head offers lower leakage performance over the MLCi, without compromise on the existing clinical merits of that design. □ Amorphous Silicon panel for iViewGT™ - Retractable arm for iViewGT™ - Software license for the iViewGT™ portal imaging system - PC for acquisition of amorphous silicon images □ Field size - The radiation field size is continuously variable from 0.5 cm × 0.5 cm to 40 cm × 40 cm in the plane con |
| 장비이미지코드 | http://www.zeus.go.kr/storage/images//equip/photo/201602/201602259056768.png |
| 장비위치주소 | 한국표준과학연구원 화학동 |
| NFEC 등록번호 | NFEC-2016-02-208017 |