시설장비 설명 1. Pure water
1) Production : Type-II grade water by Electronic driven deionization system or Reverse Osmosis system
2) Production Rate : 10 liter/hour or better
- Steady flow rate regardless of feed water temperature.
3) Pretreatment PAK cartridge
- Capacity : 30000 liter
- This cartridge should be a unique and easy-to-install pack unit(plug-and-use concept) includes three types of purification media as below
ㅇ Activated carbon
ㅇ Antiscaling compound - Polyphosphate
ㅇ 1㎛ pre-filter
- This cartridges have a built-in RFID tag to avoid maitenance errors and to improve
traceability.
4) Output product quality
- Pure Water Resistivity at 25℃ : > 5㏁.㎝
- Microorganisms(cfu/ml) with Millipak or Biopak end filter : <1cfu/㎖
- Particulates <0.22㎛(/mL) with Millipak or Biopak end filter: <1
- Pyrogen Levels with Biopak end filter : <0.001 EU/㎖
- TOC : <30ppb
- RNase Levels with Biopak end filter : <0.01ng/㎖
- DNase Levels with Biopak end filter : 99% retention
* RO membrane cut-off size : 100MW
** Upgradeable regard to Product water Quality & Quantity by the addition of
internal materials
5) Ro high recovery with Ro reject circulation:
- Low Water consumption extend PAK life
- % recovery : 24%
- Permeate water : 10 (L/H)
- Reject water : 41 (L/H)
6) Ultra-violet(UV) module
- A single-wavelength UV lamp(254nm) must be included in the system to
ensure keep bacteria levels to a minimum at the same time.
- A Log reduction value (LRV) of 4 in the bacterial count of the water
- Automatic system display alerts for UV lamp exchange
- The design of the lamp/ballast assembly enable the lamp operate efficiently in regular conditions for 2yeas without replacement.
- Clean simple lamp housing design
: Easy to change UV lamp
: UV light does not contact tubingfitting or O-rings
7) Alphanumeric display in LCD
- Flush
- Feed water conductivity (Cell constant 0.35㎝-1)with temperature compensation
- RO Product conductivity (Cell constant 0.35㎝-1)with temperature compensation
- Product conductivity (Cell constant 0.01㎝-1)with temperature compensation
- % rejection
- Temperature of water(±0.1℃)
- Stand by
- Mark Tank level in every 5% (0%~100%)
- Number of days remaining before Progard exchange
- Change UV lamp
- Progard Pack not it place
- Exchange pack
- Rinsing
- Cl2 cleaning
- pH cleaning
- RO membrane operating pressure
- Low pressure
- Battery
- Error Number : 15 codes
8) This system should check relevant parameters after each purification step
- Feed pressurefeed water quality
- RO pressureRO water quality and RO membrane efficiency(% rejection of ions)
- Elix system water quality and temperature
(using the same accurate coaxial resistivity cell as the one used in Milli-Q ultrapure water system)
9) Built-in automatic flush fuction : allows high flow of water to pass tangentially
across the surface of the RO membrane.
10) Programmed auto-clean cycle help reverse osmosis membrane is self-maintaining
11) 3-way valve in permeate line enables RO to make reliable & high quality RO water
by rejecting initial permeate water.
- Rejecting time : 3 min or more
- Rejecting water quality : 40 us/cm conductivity or more (programmable)
12) 3 step cleaning cycle extend RO life time
- Cl2 cleaning
- pH cleaning
- RO rinse
13) This system can be upgraded to a higher flow rate
14) Comply GLP standard store oprating parameters for prior 6 month and can print
out through RS232
2. Ultrapure water
1) Ultrapure water quality
- Nominal Product rates : up to 2.0 l/min
- Resistivity : 18.2 M ohm cm w/ temp compensate
- Pyrogen level with Biopak end filter : 0.001 Eu/ml.
6 log reduction with feed water >10000 Eu/ml
- Rnase level with Biopak end filter : <0.01ng/ml
- Dnase level with Biopak end filter : 0.22 um < 1/ml
- Total dissolved solids < 5ppb (if TOC ≤ 5ppb)
- Silicates < 1 ppb
- Heavy metals < 0.05ppb
- Microorganisms < 1 cfu/ml
- Ion<0.5 ppb
장비이미지코드 http://nfec.ntis.go.kr/storage/images/equip/photo/201108/.thumb/20110830095034.jpg
장비위치주소 부산 기장군 일광면 횡계리 27번지 해양생물산업육성센터 기업지원동 3층 306
NFEC 등록번호 NFEC-2010-12-129718
예약방법
카타로그 URL
메뉴얼 URL
원문 URL http://www.zeus.go.kr/equip/read?equipId=Z-NTIS-0018157
첨부파일

추가정보

과학기술표준분류, ICT 기술분류, 주제어 순으로 구성된 표입니다.
과학기술표준분류
ICT 기술분류
주제어 (키워드)