장비 및 시설 기본정보

고성능액체크로마토그래피

장비 개요

기관명, 장비번호, 제작사, 모델명, 장비사양, 취득일자, 취득금액 순으로 구성된 표입니다.
기관명 ZEUS
장비번호
제작사 Agilent Technologies
모델명 1260
장비사양
취득일자 2011-03-29
취득금액

보유기관 및 이용정보

보유기관명, 보유기관코드, 활용범위, 활용상태, 표준코드, 표준분류명, 시설장비 설명, 장비이미지코드, 장비위치주소, NFEC 등록번호, 예약방법, 카타로그 URL, 메뉴얼 URL, 원문 URL, 첨부파일 순으로 구성된 표입니다.
보유기관명 진주바이오산업진흥원
보유기관코드
활용범위
활용상태
표준코드 B402
표준분류명
시설장비 설명 1. For the analysis of wide ranges of organic compounds.
2. Single pump low pressure 4 gradient system provides automatic
system washing and equilibration as well as gradient capability.
3. Self diagnosis capability including leak detection ensures accurate detection &
treatment in case of system error.
4. Start/stop operation each system component is synchronized to ensure system safety
in case of ystem failures.
5. Memory protection against power failure of each component module is provided.
6. Remote diagnostics allow service for remote trouble shooting without
requiring an onsite visit.
7. System should supply to validation binder.
8. Automated hardware validation and system suitability test for GLP GMP cGMP.
4- Gradient Pump
4-Channel on-line vacuum degasser
Automatic liquid sampler
Column oven
Diode array UV/Vis Detector
System S/W with PC
Deuterium 램프(UV 영역 : 190-450nm)와 텅스텐 램프(Vis 영역 : 450-950nm)
190에서 950 nm 까지 전 파장 영역에서 최고의 감도를 보장
System Configuration
1. Quaternary gradient solvent delivery system 1 set
2. Automatic Liquid Sampler 1 set
3. Thermostatted Column Compartment 1 set
4. Diode Array UV/Vis Detector 1 set
5. Data & System Control Station 1 set
Specification
1. Gradient module with solvent delivery
1) Type
Dual piston with servo-controlled variable stroke drive floating
pistons and active inlet valve.
2) Flow range
0.001-10.00 ml/min or wider in 0.001 ml/min or less increments.
3) Automatic piston seal cleaning by water circulating.
4) Piston displacement
20 to 100 μl or wider automatically matched flow rate or
user-selectable.
5) Flow precision
Typically less than 0.15% based on retention time.
6) Pressure range
0-400 bar (5880 psi) or wider up to 5 ml/min or higher 0-200 bar or
wider up to 10 ml/min or higher.
7) Pressure pulsation
Typically less than 1% 1 ml/min propanol at all pressure higher than
10 psi.
8) Compressibility compensation
User-selectable based on mobile phase compressibility.
9) Gradient formation
Low pressure quaternary mixing/gradient capability using high-speed
proportioning valve. Delay volume 150 to 1500μl or less.
10) Composition range
0 to 100% in 0.1% increments or less from four independent channels.
11) Composition precision
Typically +/- 0.15% or better.
12) Safety features
a. Extensive diagnostics error detection and display via front-panel
LEDs and status logbook.
b. User definable shutdown method activated in case of error.
c. Leak detection and safe leak handling.
d. Column pressure protection with maximum rate of pressure change of
20 bar/sec or less after a set point change.
13) On-line Vacuum degasser
a. Maximum Flow rate : 10ml/min per channel or more
b. No. of Channels : 4 or more
c. Internal Volume : 1ml or less
d. Materials in contact with solvent : PTFE or PEEK
3. Automatic Liquid Sampler
1) Injection range
Programmable from 0.1 to 100 ㎕ or wider
2) Replicate injections: 1-99 or more from one vial.
3) Precision : Typically less than 0.5% RSD
4) Minimum sample volume
1 μl can be sampled from 5ul or less. then 1% 1-5ul
5) Carryover: Typically less than 0.2%.
6) Sample capacity: 96 vials or more .
7) Automated pre-column derivatization capability is provided.
8) Random Vial Access
9) Priority position for analysis of urgent samples
10) Add vials during automated run
11) Add vials to existing sequence without interruption
12) Vial Size : 2ml 5ml or more
4. Thermostatted Column Compartment
1) Temperature range: ambient to 80℃ or wider
2) Temperature stability: ± 0.15℃ or less
3) Temperature accuracy: ± 0.5℃ or less with calibration
4) Column capacity: Three or more 30-cm column
5) GLP feature : Column-identification module for GLP documentation
6) Column switching valve : Auto-filtration Concentration.
5. Diode-Array UV-Visible Detector
1) Type
UV lamp des
장비이미지코드 http://nfec.ntis.go.kr/storage/images/equip/photo/201203/.thumb/20120321134204.JPG
장비위치주소 경남 진주시 문산읍 삼곡리 월아산로 991 재단법인바이오21센터 재단법인바이오21센터 시험생산동 2층 207
NFEC 등록번호 NFEC-2012-03-157718
예약방법
카타로그 URL
메뉴얼 URL
원문 URL http://www.zeus.go.kr/equip/read?equipId=Z-NTIS-0031531
첨부파일

추가정보

과학기술표준분류, ICT 기술분류, 주제어 순으로 구성된 표입니다.
과학기술표준분류
ICT 기술분류
주제어 (키워드)