전자선형가속기
기관명 | ZEUS |
---|---|
장비번호 | |
제작사 | Elekta |
모델명 | Synergy® Platform |
장비사양 | |
취득일자 | 2016-02-24 |
취득금액 |
보유기관명 | 한국표준과학연구원 |
---|---|
보유기관코드 | |
활용범위 | |
활용상태 | |
표준코드 | F0 |
표준분류명 | |
시설장비 설명 | □ Reproducibility - Maximum coefficient of variation of the ratio of number of dose monitor units and absorbed dose is ≤0.5%. This is applicable for any dose rate and to every individual segment of all conformal and intensity modulated delivery techniques. □ Proportionality (linearity) - The maximum deviation of measured absorbed dose by an external dose meter at the isocenter from the number of delivered monitor units multiplied by the proportionality (calibration) factor is ≤1% or 0.1 MU, whichever is greater for 1 to 999 MU. □ MLCi2 Head - Lower Leakage Precision Multileaf Collimator - The MLCi2 Head offers lower leakage performance over the MLCi, without compromise on the existing clinical merits of that design. □ Amorphous Silicon panel for iViewGT™ - Retractable arm for iViewGT™ - Software license for the iViewGT™ portal imaging system - PC for acquisition of amorphous silicon images □ Field size - The radiation field size is continuously variable from 0.5 cm × 0.5 cm to 40 cm × 40 cm in the plane con |
장비이미지코드 | http://www.zeus.go.kr/storage/images//equip/photo/201602/201602259056768.png |
장비위치주소 | 한국표준과학연구원 화학동 |
NFEC 등록번호 | NFEC-2016-02-208017 |
예약방법 | |
카타로그 URL | |
메뉴얼 URL | |
원문 URL | http://www.zeus.go.kr/resv/equip/read/Z-NTIS-0060100 |
첨부파일 |
과학기술표준분류 | |
---|---|
ICT 기술분류 | |
주제어 (키워드) |