1. Spectrometer must be provided better than minimum of 0.125 ㎝-1 resolution for without potion and be step selectable to 128 ㎝-1 resolution and expandable spectral range from 2500~ 10 ㎝-1 with automated dual source detector and three beamsplitters combinations. 2. Spectrometer must be capable of computer controlled selection of multiple spectral ranges without the need to manually change beamsplitters. Automatically change of spectral ranges to include beamsplitter source and detector configurations should be accomplished by soft ware control.1. spectral range : 25000~50 ㎝-1 2. standard resolution : step selectable 128.0~0.125 ㎝-1 3. wave number precision : better than 0.01 ㎝-1 using He-Ne laser in quadrature 4. photometric precision : better than ±0.07T at all wavelengths 5. Interferometer : high-throughput 60。 Michelson with corner-cube retroreflectors.분자진동 모드 조사 |