IN-SITU OBJECT COUNTING SYSTEM
기관명 | ZEUS |
---|---|
장비번호 | |
제작사 | Canberra Industries(U.S.A) |
모델명 | ISOCS |
장비사양 | |
취득일자 | 2004-05-12 |
취득금액 |
보유기관명 | 한국원자력연구원 |
---|---|
보유기관코드 | |
활용범위 | |
활용상태 | |
표준코드 | |
표준분류명 | |
시설장비 설명 | 특징 우라늄변환시설 해체공사시 변환공정 기기 탱크 펌프 배관 등에서 시료를 채취하지 않고 현장에서 직접 대상물의 방사선오염 핵종 및 방사능농도 측정구성및성능 1. Coxial Germanium Detector * Efficiency:40% * Effective energy range:40keV to 10MeV * Resolution(FWHM):Less than 1.0keV at 122keV 1.9keV at 1332keV 2.Flanged Big MAC cryostat 3.Remote Drtector Chamber(4"neck) 4.InSpector 2000 DSP Portable MCA * Fully integrated portabe MCA including Digital Signal Processor high voltage power supply digital stabilizer and 16K memory/conversion gain * Provided with softside carring case battery & charger AC adaptor all-in-one detector cable communication cables(USB&RS-232) and G2K InSpector Basic Spectroscopy Software(S504C) 5.Genie-2000 Gamma Option * Peak area background subtract efficiency correction nuclide identification automatic interference correction weighted mean activity calculation and minimum detectable activity 6.Genie-2000 Quality Assurance * Monitors performance of the spectroscopy system over time * Any parameter in the system can be tracked and analyzed * Two level alarm limits 7.Genie-2000 Interactive Peak Fit * Allows user to review wesults of peak search add and delete peaks or modify fit regions * Graphical display of fit and residuals * Requires ISOXCAL 8.ISOCS In-Situ Efficiency Calibration Software * Mathematical efficiency calibration software for Ge detectors * No radioactive sources needed * Includes 20 different templates for wide range of source-detector geometries * Requires ISOXCAL 9.ISOCS Characterization for New CI Detector * Requires precious purchase of ISOCS software 10.Verification Counts for ISOCS * Validation counts and report of characteriwed detector * Uses 4 typical laboratory standards(Marinelli beaker filter paper 350cc 20cc) * 7different calibration geometrics modeled with ISOCS 11. Wheenull |
장비이미지코드 | http://nfec.ntis.go.kr/storage/images/equip/photo/201104/.thumb/20110408131121.jpg |
장비위치주소 | 대전 유성구 덕진동 150 한국원자력연구원 |
NFEC 등록번호 | NFEC-2005-11-044850 |
예약방법 | |
카타로그 URL | |
메뉴얼 URL | |
원문 URL | http://www.zeus.go.kr/equip/read?equipId=Z-NTIS-0016273 |
첨부파일 |
과학기술표준분류 | |
---|---|
ICT 기술분류 | |
주제어 (키워드) |