시설장비 설명 |
- Measurement Method : 680℃ Combustion catalystic oxidation/non-dispersive infrared gas analysis method - Operation Method : PC-controlled - Measuring Items : TC IC TOC NPOC (Optional POC TN) - Appliable sample : aqueous sample - Measuring range : TC- 0 to 25000mg/L IC- 0 to 30000mg/L - Detection limit : 4ug/L - Measurement accuracy : CV 1.5% Max - Measurement time : Approx. 3~4 minutes for both TC and IC - Sample injection : Automatic injection via microliter syringe - Sample injection volume : 10 to 2000uL variable - IC Pre-treatment : Automatic Internal acidification and aparging - Automatic diluation : dilution factor 2 to 50 - Gas consumption : approx. 1440L/month - Data Processing Detector Signal: Linearization baseline correction peak detection and area calculation - |