시설장비 설명 |
1. Features 1) The analyzer must provide the high degree of accuracy in determining Total Carbon(TC), Total Organic Carbon(TOC), Non-purgeable Organic Carbon(NPOC) and Inorganic Carbon(IC) content in drinking water and wastewater samples. 2) The analyzer must use a combustion method technology with a high temperature furnace to oxidize organic compounds. 3) A Non-Dispersive Infrared(NDIR) detector will accomplish detection of the carbon dioxide resulting from purging of Inorganic Carbon and oxidation of Organic Carbon. 4) Precise and adjustable sample injection system for automatic analysis(5) Autosampler in conjunction with analyzer, the unit must be able to pre-acdify then sparge the sample in vial to remove inorganic carbon and must have septum piercing function for capped vials contain samples. 6) User can transfer past or current sample results, statistical and other sample information into spread sheet, like Microsoft EXCELTM and the transfer sample results and statistcal information into spreadsheet ca |